Bredow, J., Beyer, F., Oppermann, J., Kraus, B., Meyer, C., Stein, G., Eysel, P. and Koy, T. (2016). A novel classification of screw placement accuracy in the cervical spine. Technol. Health Care, 24 (6). S. 919 - 926. AMSTERDAM: IOS PRESS. ISSN 1878-7401

Full text not available from this repository.

Abstract

BACKGROUND: Goal of this study is to present an easily reproducible and reliable measurement to evaluate accuracy of screw placement in cervical spine. METHODS: Accuracy of cervical screw position was assessed in 52 patients treated with 163 screws. Each patient receiving pedicle, transarticular C1/2, lateral mass, or laminar screws and postoperative CT scan was included. Placement position was categorized in 5 grades: Grade 1 is ideal with pedicle wall perforation < 1 mm, grade 2 < 2 mm, grade 3 < 3 mm, and grade 4 < 4 mm. Grade 5 > 4 mm and/or obstruction of transverse foramen by more than half a screw diameter. Intraclass correlation coefficient (ICC) values were assessed for inter-and intraobserver reliability. RESULTS: The mode of individual evaluations was calculated to assign a single value to each screw. This yielded 89 grade 1 (54.6%), 48 grade 2 (29.4%), 14 grade 3 (8.6%), 3 grade 4 (1.8%), and 9 grade 5 (5.5%) screws. Intraobserver reliability ICC was 0.966 and 0.959 for measurements. Interobserver reliability ICC was 0.938. CONCLUSION: This study introduces a reliable classification of cervical spine instrumentation with various screw types. This should enable the use of a uniform and reproducible, and thus comparable classification for screw position in cervical spine.

Item Type: Journal Article
Creators:
CreatorsEmailORCIDORCID Put Code
Bredow, J.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Beyer, F.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Oppermann, J.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Kraus, B.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Meyer, C.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Stein, G.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Eysel, P.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
Koy, T.UNSPECIFIEDUNSPECIFIEDUNSPECIFIED
URN: urn:nbn:de:hbz:38-288573
DOI: 10.3233/THC-161246
Journal or Publication Title: Technol. Health Care
Volume: 24
Number: 6
Page Range: S. 919 - 926
Date: 2016
Publisher: IOS PRESS
Place of Publication: AMSTERDAM
ISSN: 1878-7401
Language: English
Faculty: Unspecified
Divisions: Unspecified
Subjects: no entry
Uncontrolled Keywords:
KeywordsLanguage
PEDICLE SCREW; LATERAL MASS; NONTRAUMATIC LESIONS; PULLOUT STRENGTH; FIXATION; FLUOROSCOPY; NAVIGATION; COMPLICATIONS; STABILITY; GUIDANCEMultiple languages
Health Care Sciences & Services; Engineering, BiomedicalMultiple languages
Refereed: Yes
URI: http://kups.ub.uni-koeln.de/id/eprint/28857

Downloads

Downloads per month over past year

Altmetric

Export

Actions (login required)

View Item View Item