Experimental Comparison of Fiducial Markers Used in Proton Therapy: Study of Different Imaging Modalities and Proton Fluence Perturbations Measured With CMOS Pixel Sensors

Reidel, Claire-Anne ; Horst, Felix ; Schuy, Christoph ; Jäkel, Oliver ; Ecker, Swantje ; Henkner, Katrin ; Brons, Stephan ; Durante, Marco ; Weber, Uli (2022)
Experimental Comparison of Fiducial Markers Used in Proton Therapy: Study of Different Imaging Modalities and Proton Fluence Perturbations Measured With CMOS Pixel Sensors.
In: Frontiers in Oncology, 2022, 12
doi: 10.26083/tuprints-00021125
Article, Secondary publication, Publisher's Version

[img] Text
fonc-12-830080.pdf
Copyright Information: CC BY 4.0 International - Creative Commons, Attribution.

Download (6MB)
Item Type: Article
Type of entry: Secondary publication
Title: Experimental Comparison of Fiducial Markers Used in Proton Therapy: Study of Different Imaging Modalities and Proton Fluence Perturbations Measured With CMOS Pixel Sensors
Language: English
Date: 9 May 2022
Place of Publication: Darmstadt
Year of primary publication: 2022
Publisher: Frontiers Media S.A.
Journal or Publication Title: Frontiers in Oncology
Volume of the journal: 12
Collation: 18 Seiten
DOI: 10.26083/tuprints-00021125
Corresponding Links:
Origin: Secondary publication DeepGreen
Abstract:

Fiducial markers are used for image guidance to verify the correct positioning of the target for the case of tumors that can suffer interfractional motion during proton therapy. The markers should be visible on daily imaging, but at the same time, they should produce minimal streak artifacts in the CT scans for treatment planning and induce only slight dose perturbations during particle therapy. In this work, these three criteria were experimentally investigated at the Heidelberg Ion Beam Therapy Center. Several small fiducial markers with different geometries and materials (gold, platinum, and carbon-coated ZrO₂) were evaluated. The streak artifacts on treatment planning CT were measured with and without iMAR correction, showing significantly smaller artifacts from markers lighter than 6 mg and a clear improvement with iMAR correction. Daily imaging as X-ray projections and in-room mobile CT were also performed. Markers heavier than 6 mg showed a better contrast in the X-ray projections, whereas on the images from the in-room mobile CT, all markers were clearly visible. In the other part of this work, fluence perturbations of proton beams were measured for the same markers by using a tracker system of several high spatial resolution CMOS pixel sensors. The measurements were performed for single-energy beams, as well as for a spread-out Bragg peak. Three-dimensional fluence distributions were computed after reconstructing all particle trajectories. These measurements clearly showed that the ZrO₂ markers and the low-mass gold/platinum markers (0.35mm diameter) induce perturbations being 2–3 times lower than the heavier gold or platinum markers of 0.5mm diameter. Monte Carlo simulations, using the FLUKA code, were used to compute dose distributions and showed good agreement with the experimental data after adjusting the phase space of the simulated proton beam compared to the experimental beam.

Uncontrolled Keywords: fiducial marker, dose perturbation, proton therapy, CMOS pixel sensor, image guidance, streak artifacts, Monte Carlo simulation
Status: Publisher's Version
URN: urn:nbn:de:tuda-tuprints-211254
Classification DDC: 500 Science and mathematics > 530 Physics
500 Science and mathematics > 570 Life sciences, biology
600 Technology, medicine, applied sciences > 610 Medicine and health
Divisions: 05 Department of Physics > Institute for Condensed Matter Physics > Biophysics
Date Deposited: 09 May 2022 13:37
Last Modified: 14 Nov 2023 19:04
SWORD Depositor: Deep Green
URI: https://tuprints.ulb.tu-darmstadt.de/id/eprint/21125
PPN: 499795679
Export:
Actions (login required)
View Item View Item